描述
CYROLITE G-20 CP UV Protective Medical Acrylic Copolymer
Material: PMMA-based Copolymer
UV Transmission: <1% (260-480nm)
Impact: 5x Acrylic
Medical: USP Class VI
Packing: 1500 lb boxes
CYROLITE G-20 CP provides <1.0% UV transmission (260-480nm) to protect light-sensitive drugs while maintaining excellent visible light transmission for clear visibility of fluid pathways.
CYROLITE G-20 CP is a UV protective, medical grade PMMA-based copolymer designed specifically for infusion therapy medical devices requiring protection of light-sensitive pharmaceutical compounds. With UV light transmission of less than 1.0% in the 260-480nm range, this material effectively shields oncology drugs, antibiotics, and antifungal agents from photodegradation during administration. Simultaneously, it maintains excellent visible light transmission between 500-780nm, ensuring clear visibility of fluid pathways and patient safety.
CYROLITE G-20 CP combines the proven properties of the CYROLITE G-20 platform — 5 times the impact resistance of unmodified acrylics, excellent bonding to PVC tubing, compatibility with EtO, gamma, and E-beam sterilization — with specialized UV protection for photosensitive drug delivery. It meets USP Class VI and ISO 10993-1 biocompatibility requirements, making it suitable for patient-contact medical devices. Processable on standard injection molding and extrusion equipment with 3-zone general purpose screws. Typical applications include filter housings, Y-sites, luer connectors, needleless connectors, and check valves for infusion therapy and oncology drug delivery.
Genuine Roehm CYROLITE Imported Stock and Technical Support
We supply stable spot inventory of genuine CYROLITE G-20 CP UV protective medical grade acrylic copolymer. Official TDS, batch COA, and biocompatibility documentation (USP Class VI, ISO 10993) are available for infusion therapy and medical device applications.
Request CYROLITE G-20 CP Quotation, TDS, and Sample
Complete TDS, COA, and regulatory documentation available
Core Advantages of CYROLITE G-20 CP
- UV light protection — <1.0% transmission in 260-480nm range for photosensitive drugs
- Excellent visible clarity — maintains visibility of fluid pathways (500-780nm)
- Medical grade compliance — USP Class VI and ISO 10993-1 certified
- High impact resistance — 5 times the impact resistance of unmodified acrylics
- Multiple sterilization compatibility — EtO, gamma, and E-beam
- Chemical and body fluid resistance — suitable for drug delivery applications
- Excellent bonding to PVC — thermal, ultrasonic, and laser weldable
- Versatile processing — injection molding and extrusion
Main Applications
- Filter housings — UV-protective housings for drug infusion filters
- Y-sites — medication injection sites for oncology drugs
- Luer connectors — UV-protective fluid pathway connections
- Needleless connectors — safe access ports for light-sensitive medications
- Check valves — one-way fluid control for infusion therapy
- Infusion therapy devices — oncology, antibiotic, and antifungal drug delivery
Full Technical Parameters
| Property | Typical Value | Test Method |
| Material Type | PMMA-based copolymer | — |
| Medical Compliance | USP Class VI, ISO 10993-1 | — |
| UV Transmission (260-480nm) | <1.0% | ASTM D1003 |
| Melt Flow Rate (230°C/5.0kg) | 2.6 g/10min | ASTM D1238 |
| Tensile Strength | 46.9 MPa | ASTM D638 |
| Tensile Modulus | 2.2 GPa | ASTM D638 |
| Tensile Elongation @ Break | 9.5% | ASTM D638 |
| Flexural Strength | 72.4 MPa | ASTM D790 |
| Flexural Modulus | 2.3 GPa | ASTM D790 |
| Notched Izod Impact (23°C) | 101 J/m | ASTM D256 |
| Vicat Softening Point | 101°C | ASTM D1525 |
| HDT (1.8 MPa) | 86°C | ASTM D648 |
| Specific Gravity | 1.11 | ASTM D792 |
| Mold Shrinkage | 0.004-0.007 in/in | ASTM D955 |
| Sterilization | EtO, Gamma, E-beam | — |
Note: All listed figures are typical values and may vary by production batch. Complete independent qualification testing is required before mass production deployment.
Recommended Processing Conditions
| Parameter | Recommended Value |
| Processing Method | Injection Molding / Extrusion |
| Screw Type | 3-zone general purpose screw |
| Melt Temperature | 230-250°C (typical) |
| Mold Temperature | 50-80°C (typical) |
Handling and Storage Recommendations
| Storage Rules | Store in dry, cool conditions away from direct sunlight and heat sources. |
| Packaging | 1500 lb gaylor boxes; other packaging available on request. |
| Operation Protection | Follow standard safety practices for handling thermoplastic pellets. Use appropriate PPE during processing. |
Official Technical Documentation
| Document Type | Document Introduction | Document Request |
| TDS | CYROLITE G-20 CP original factory technical data sheet | Request TDS |
| COA | Batch-specific certificate of analysis | Request COA |
| Biocompatibility | USP Class VI and ISO 10993-1 documentation | Request Documentation |
Frequently Asked Questions
What is CYROLITE G-20 CP?
CYROLITE G-20 CP is a UV protective medical grade acrylic copolymer designed to shield light-sensitive drugs (oncology, antibiotics, antifungals) from degradation during infusion therapy. It provides <1% UV transmission (260-480nm) while maintaining excellent visible light transmission.
What is the UV transmission of CYROLITE G-20 CP?
CYROLITE G-20 CP provides <1.0% UV light transmission in the 260-480nm range, effectively protecting photosensitive pharmaceutical compounds while maintaining excellent visibility in the 500-780nm range.
What medical certifications does CYROLITE G-20 CP hold?
CYROLITE G-20 CP meets USP Class VI and ISO 10993-1 biocompatibility requirements, making it suitable for medical device applications requiring patient contact.
What applications is CYROLITE G-20 CP suitable for?
CYROLITE G-20 CP is ideal for infusion therapy medical devices requiring UV protection, including filter housings, Y-sites, luer connectors, needleless connectors, and check valves for light-sensitive drug delivery.
What sterilization methods are compatible with CYROLITE G-20 CP?
CYROLITE G-20 CP is resistant to EtO, gamma, and E-beam sterilization, providing flexibility for medical device manufacturing.
Is CYROLITE G-20 CP bondable to PVC tubing?
Yes. CYROLITE G-20 CP offers excellent bonding to PVC tubing and can be thermal bonded, ultrasonic welded, and laser welded.
What is the impact resistance of CYROLITE G-20 CP?
CYROLITE G-20 CP offers 5 times the impact resistance of unmodified acrylics, with a notched Izod impact strength of 101 J/m at 23°C.
Related CYROLITE Products
Compare CYROLITE medical grade polymers for different application requirements.
| Product | Key Feature | MFR | USP Class VI | Core Application |
| G-20 CP | UV Protection (<1% 260-480nm) | 2.6 | ✅ | Light-sensitive drug delivery |
| G-20 HIFLO | High flow (10.0 MFR) | 10.0 | 001 tint | Thin-wall components |
| CYROLITE G-20 | General medical grade | 2.6 | colorless | General medical devices |